Visante® anterior segment OCT in a patient with gas bubbles in the anterior chamber after femtosecond laser corneal flap formation


YILDIRIM C. A., Altunsoy M., Basar D.

International Ophthalmology, cilt.30, sa.1, ss.81-84, 2010 (Scopus) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 30 Sayı: 1
  • Basım Tarihi: 2010
  • Doi Numarası: 10.1007/s10792-008-9276-5
  • Dergi Adı: International Ophthalmology
  • Derginin Tarandığı İndeksler: Scopus
  • Sayfa Sayıları: ss.81-84
  • Anahtar Kelimeler: Air bubbles, Femtosecond laser, IntraLase, LASIK, Visante
  • Dokuz Eylül Üniversitesi Adresli: Evet

Özet

A patient with anterior chamber gas bubbles after LASIK flap formation with femtosecond laser is presented. A 33-year-old male patient had LASIK for myopia with corneal flap formation with IntraLase® FS30 femtosecond laser. In the right eye, air bubbles were observed in anterior chamber after successful flap formation. Laser correction was completed successfully, by Wavelight® Allegretto WaveTM Eye-Q excimer laser, without eye-tracker. Visante® anterior segment optical coherence tomography imaging displayed that the pocket extended to limbal area, with stromal bed thickness of >600 μm at that area. We propose as possible causes of the air bubbles scattering of femtosecond laser beam producing direct photodissociation of aqueous humor, or direct cavitation effect in aqueous humor due to rapid pressure changes in corneal lamellar interface during the femtosecond laser procedure. Other causes such as migration of intrastromal bubbles through peripheral corneal stroma and trabecular meshwork, or through posterior stroma and endothelium, seem to be unlikely © 2008 Springer Science+Business Media B.V.