Optical determination of glucose with glucose oxidase immobilized in PVC together with fluorescent oxazol-5-one derivatives


Ozturk G., TİMUR S., ALP S.

ANALYTICAL LETTERS, cilt.41, sa.4, ss.608-618, 2008 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 41 Sayı: 4
  • Basım Tarihi: 2008
  • Doi Numarası: 10.1080/00032710801910551
  • Dergi Adı: ANALYTICAL LETTERS
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.608-618
  • Anahtar Kelimeler: glucose oxidase, glucose, fluorescence emission, polyvinyl chloride (PVC) films, OXYGEN OPTRODE, BIOSENSOR, TRANSDUCER, SENSOR, NMR
  • Dokuz Eylül Üniversitesi Adresli: Evet

Özet

Fully reversible biosensors for glucose monitoring based on solid polyvinylchloride (PVC) films where the enzyme glucose oxidase (GOx) was incorporated together with 2-phenyl-4-[4-(1,4,7,10-tetraoxa-13-azacyclopentadec yl)benzylidene]-5-oxazolone (CPO-1), 2-(3,5-dinitrophenyl)-4-[4-(1,4,7,10- tetraoxa-13-azacyclopentadecyl)benzylidene]-5-oxazolone (CPO-2), 2-( 4- nitrophenyl) -4-[ 4-( 1,4,7,10-tetraoxa-13-azacyclopentadecyl)benzylidene]-5-oxazolone (CPO-3) and 2-(4-tolyl)-4-[4-(1,4,7,10-tetraoxa-13-azacyclopentadecyl)benzylidene]-5-oxazolone (CPO-4) derivatives have been developed. The limit of detection (LOD) values for glucose were found to be 1.47 x 10 25 M, 2.01 x 10 25 M, 0.89 x 10 25 M, and 0.12 x 10 25 M for CPO-1, CPO-2, CPO-3, and CPO-4, respectively (n = 7). Sensor films were found to have excellent photostability.