Computational analysis of the effects of interprosthetic distance on normal and reduced cortical thickness femur models


Creative Commons License

Gungorurler M., Gürsan O., Havıtçıoğlu H.

PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2021 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası:
  • Basım Tarihi: 2021
  • Doi Numarası: 10.1177/09544119211040998
  • Dergi Adı: PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Aerospace Database, Applied Science & Technology Source, Biotechnology Research Abstracts, CINAHL, Communication Abstracts, Compendex, EMBASE, INSPEC, MEDLINE, Metadex, Civil Engineering Abstracts
  • Anahtar Kelimeler: Finite element (biomechanics), interprosthetic fracture, cortical strain, strain shielding, hip biomechanics, TOTAL HIP-ARTHROPLASTY, BONE-MINERAL DENSITY, FINITE-ELEMENT, FRACTURE FIXATION, STRESS, STEM, PRINCIPLES
  • Dokuz Eylül Üniversitesi Adresli: Evet

Özet

Distal femoral fractures associated with the femoral stem in a well-fixed hip arthroplasty pose a risk of an interprosthetic fracture, the treatment of which is known as difficult. To effectively prevent and treat IP fractures, biomechanical effects must be demonstrated. We defined eight variations of the interprosthetic distance ranging from 48 mm overlap to 128 mm gap. Femoral geometries with normal and reduced cortical thickness were modeled to evaluate the effects of cortical thickness. In addition to the intact model, a total of 16 finite element models were analyzed under physiological boundary conditions. Maximum and minimum principal strains on the lateral and medial cortex surfaces were always found to be greater in models with reduced cortical thickness than in normal femurs. The model with 48 mm overlapping interprosthetic distance produced the least peak strain and the model with 16 mm interprosthetic gap produced the greatest strain with both normal and reduced cortical thickness. The screw holes produced local strain concentrations and increased the peak strains on the cortex surfaces, especially close to the stem tip. Statistically, a significant correlation (R-2 = 0.9483) was found between strain shielding and interprosthetic distance. Axial stiffness, interfragmentary shear motion, and maximum von-Mises stress on the distal plate showed a high correlation with the interprosthetic distance. It was concluded that the overlapping structures are superior to other fixations we analyzed in that they offer better mechanical stability and eliminates the local strain concentrations.