A study on the extrinsic sensitivity and counting, efficiency of a gamma camera for a cylindrical source and a rectangular detector


Haliloglu R. C., Karadeniz Ö., Durak H.

APPLIED RADIATION AND ISOTOPES, cilt.130, ss.218-223, 2017 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 130
  • Basım Tarihi: 2017
  • Doi Numarası: 10.1016/j.apradiso.2017.10.001
  • Dergi Adı: APPLIED RADIATION AND ISOTOPES
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.218-223
  • Anahtar Kelimeler: Extrinsic sensitivity, Counting efficiency, Scattered radiation effects, Gamma camera, Cylindrical source, Rectangular detector
  • Dokuz Eylül Üniversitesi Adresli: Evet

Özet

In this work, the extrinsic counting efficiency and extrinsic sensitivity of a cylindrical source and a gamma camera with a rectangular detector were determined using homogeneous Tc-99(m). Scattered radiation effects were evaluated by both analyzing the energy spectrum of Tc-99m for the scatter fraction and plotting the extrinsic counting efficiency and sensitivity. It is found that the scattered fraction values increased with increasing source thickness. Calculated extrinsic sensitivity shows a gradually decreasing trend with increasing source thickness. The extrinsic efficiency decreases with increasing source thickness, It is concluded that increasing source to detector distances results with increasing extrinsic counting efficiency but decreasing extrinsic sensitivity.